Laboratory Environment Control Terminal
Ultra high precision temperature and humidity control Independent research and development of high precision temperature control technology, control output accuracy 0 . 1% The key area of the internal temperature stability of the equipment can reach/-5mk (static), the internal temperature specification is 22.oc (adjustable), and the temperature level uniformity is less than 16mk/m The internal humidity stability of the equipment can reach 0.5% @ 8h The internal pressure stability of the equipment can reach/-3pa Continuous stable working time is more than 144h |
Ultra high level of cleanliness control The system can achieve more than one hundred levels of cleanliness control. Internal cleanliness: up to ISO Class 3 (equipment work area) |
Real-time data record query Data automatically generates curves for easy viewing Data is automatically saved and can be exported in tabular form at any time Synchronous records of events such as running status and fault status, with clear query objectives |
Appearance custom detachable Removable aluminum alloy frame, large equipment can be assembled on site The box is made of high-quality sheet metal material, beautiful and generous, and the appearance color can be customized according to customer needs |
Quiet and efficient equipment operation More efficient, more powerful and quieter operation with EC fans The interior of the refrigeration unit adopts high-efficiency sound insulation material, which improves the noise of the equipment to reduce the noise <45dB |
automatic safety protection system Fault automatic protection program, all-weather worry-free operation Fault real-time sound and light alarm reminder, remote assistance fault handling |
Naming | SCIS | H | A | A | 100 |
code point | 1 | 2 | 3 | 4 | 5 |
Code 1 | SCIS-Brand | ||||
Code 2 | H-high precision ring control series | ||||
Code 3 | Refrigeration mode number A- air cooling type, refrigerant medium is water B- water cooling type, refrigerant medium is water C- air cooling type, refrigerant medium is Freon D-water cooling type, refrigerant medium is Freon E-user provided chilled water | ||||
Code 4 | Temperature accuracy No. A- Accuracy greater than ± 0.01 B- Accuracy less than ± 0.01, greater than ± 0.05 C- Accuracy less than ± 0.05 | ||||
Code 5 | Equipment scale number 100-small equipment (general refrigerating capacity less than 5KW)200-medium equipment (general refrigerating capacity between 5KV-10KW) 300-large equipment (general refrigerating capacity above 10KW) |
The equipment shall be placed in a laboratory with certain temperature and humidity accuracy and cleanliness requirements for operation. | |
Ambient temperature and humidity conditions | ± 2 ℃ DB temperature change rate within 0.5 ℃/8h |
± 5% RH humidity change rate 59% within 8h | |
Ambient cleanliness conditions | ISO class6 |
Environmental noise | ≤ 60dB |
Power supply | Three-phase five-wire system 380VAC-50Hz or 220VAC-50Hz (depending on equipment) |
Power supply voltage | Maximum allowable fluctuation range is ± 10% |
Storage ambient temperature | Equipment refrigeration unit -40 ℃-70 ℃ touch screen -20 ℃-60 ℃, no condensation |
Operating ambient temperature | Controller -20 ℃-60 ℃ Touch screen 0 ℃-50 ℃, no condensation |
Typical application
chip semiconductor
Precision measurement
Precision Manufacturing
Research and development
Immediate Consultation